Lecroy MSO64MXs-B 4-channel oscilloscope, Digital Storage oscilloscope, MSO64MXs-B Manuel D’Utilisation

Codes de produits
MSO64MXs-B
Page de 128
G
ETTING 
S
TARTED 
M
ANUAL
 
45 
 WSMXs-B-GSM-E Rev A
 
Trigger level is specified in volts and remains unchanged when you change the 
vertical gain settings or offset.  
The amplitude and range of the trigger level are limited as follows: 
 
± 4.1 screen divisions with a channel as the trigger source 
 
± 0.5 V with EXT as the trigger source 
 
± 5 V with EXT/10 as the trigger source 
 
None with LINE as the trigger source (zero crossing is used). 
Edge Trigger Set Up 
1.  Touch the trigger descriptor label at the bottom right of the display. 
2. 
 
3.  Or, touch Trigger Trigger Setup on the menu bar.  
Either way, the following dialog appears at the bottom of the display.  
 
4.  Touch inside the Source field in the Setup section and select an input from 
the pop-up menu: 
 
 
C1 through C4 are abbreviations for Channel 1 through 
Channel 4.  
 
Ext and Ext/10 allow triggering on the External Input of 
the oscilloscope. Ext input is +/-0.5V. Ext/10 input is+/-
5.0V.  
 
Line is for triggering on the Positive or Negative 
excursion of the AC power line (not available when using 
battery or DC input power). 
5.  Select a trigger coupling from the Coupling section. Coupling refers to the 
type of signal coupling at the input of the trigger circuit. You can choose from 
these coupling types: