With the Frame ID, ID + Data, or Error trigger types chosen, select either the Binary or Hexadecimal (Hex) 
setup format. The format propagates through the entire LIN trigger setup. 
Toggling back and forth between the formats does not result in lost information (binary is used internally 
as the core format for all triggering and decoding operations), though use of don’t care bits in a binary 
setup results in the display of an X (for a full nibble don’t care) or a $ (for a partial nibble don’t care). 
5.
 
F
RAME 
ID
 
S
ETUP
 
Frame ID Setup is used to trigger on a specific Frame ID value with either 11 or 29 Bits. 
When LINbus trigger selections are either Frame ID or ID + Data, use the Frame ID Setup fields as follows: 
 
ID Condition - Select from EqualNot EqualLess thanLess than or Equal toGreather thanGreater 
than or Equal to
In Range, or Out Range conditions available. 
 
Frame ID - Provide a value in either Binary or Hexadecimal mode based on the Setup Format selection 
made in the previous step. 
 
To Frame ID - If the Frame ID condition (previous) is In Range of or Out of Range, provide a value here to 
specify the full ID trigger range. 
Note: If the Frame ID is equal to 3C or 3D, the # Data Bytes field in the following Data Pattern Setup 
step defaults to 8. 
ID
 
+
 
D
ATA 
T
RIGGER 
S
ETUP 
D
ETAIL
 
6a. Data Pattern Setup 
When the LINbus trigger selection is ID + Data, use the Data Pattern Setup fields as follows:  
 
Data Value - Provide a value in either Binary or Hexadecimal mode based on the selection made in the 
previous Setup Format step. 
 
Data Value To - This field is only enabled when the following Condition field contains an In Range of or 
Out of Range value. 
PLEASE NOTE THE FOLLOWING:  
 
Up to 8 bytes of data can be entered as a pattern value. 
 
If less than 8 bytes of data is entered for the pattern value, the data is assumed to begin at Data 
Byte 1 in the LIN message. If this is not desired, then add preceding or trailing don’t care (X) nibbles 
to the pattern value. 
 
In Hexadecimal format, data must be entered as full bytes even though the minimum required 
acceptable entry is a nibble. If less than a full byte is entered, then a don’t care X precedes the 
pattern values entered. 
 
Condition - Select from EqualNot EqualLess thanLess than or Equal toGreather thanGreater than 
or Equal to
In Range, or Out Range conditions available. 
 
# Data Bytes - This field value defaults to the length, in bytes, of the pattern set in the Pattern Value 
selection. If you were to change the length to be less than this value, it would truncate the beginning of 
the pattern value. If you were to increase the pattern length, it would add don’t care XX byte values to 
the beginning of the pattern value. The maximum number of data bytes is 8, per the LIN standard.