Safety warning
We strongly recommend that you read the general safety information below before
using your product for the first time.  If the equipment is not used in the manner
specified, then the protection provided may be impaired.  This could result in damage
to your computer and injury to yourself or others.
Maximum input range
The inputs of the PicoLog 1000 Series loggers are protected against overvoltages up
to ±30 V.  Input voltages beyond this range may cause permanent damage to the
unit.
Mains voltages
Pico products are not designed for use with mains voltages.  To measure mains we
recommend the use of a differential isolating probe specifically designed for such
measurements.
Safety grounding
The ground of every unit is connected directly to the ground of your computer via the
interconnecting cable provided.  This is done to minimise interference.  If the PC
(especially laptops) is not grounded, reading stability cannot be guaranteed and it
may be necessary to manually ground the equipment.
As with most oscilloscopes and data loggers, you should take care to avoid connecting
the inputs of the unit to any equipment which may be at an unsuitable voltage.  If in
doubt, use a voltmeter to check that there is no hazardous AC or DC voltage.  Failure
to check may cause damage to the product and/or computer and could cause injury to
yourself or others.
You should assume that the product does not have a protective safety earth. Incorrect
configuration and/or use on voltages outside the maximum input range can be
hazardous.
Cleaning
The unit may be cleaned externally using a cloth dampened slightly with water.  Do
not use any other solvents or abrasive cleaning materials.
Repairs
The unit contains no user-serviceable parts.  Repair or calibration of the unit requires
specialised test equipment and may be performed only by Pico Technology or its