Basic Safety Checks 
 
Background 
The GPT-9600 Series are high voltage devices 
and as such, daily safety checks should be 
made to ensure safe operation.  
 
1.  Ensure all test leads are not broken and are free 
from defects such as cracks or splitting.  
 
2.  Ensure the safety tester is always connected to 
an earth ground. 
 
3.  Test the safety tester operation with a low 
voltage/current output:  
 
Ensure the safety tester generates a FAIL 
 
judgment when the HIGH VOLTAGE and 
 
RETURN terminals are shorted (using the 
 
lowest voltage/current as the testing 
 
parameters).  
 WARNING 
Do not use high voltages/currents when the HIGH 
VOLTAGE and RETURN terminals are shorted. It 
may result in damage to the instrument.