OPERATION 
31 
IR Grounding Mode Information 
 
Background 
For IR tests the GROUND MODE is always set 
to OFF. Here, the return terminal is floating 
with respect to the earth ground. This mode is 
for DUTs that are floating and not directly 
connected to an earth ground. In this mode any 
stray capacitance/resistance that leaks to the 
earth ground from the DUT side of the testing 
circuit and will not be measured.  
 Warning 
When GROUND MODE is set to OFF, the DUT, 
fixtures or connected instrumentation cannot be 
grounded. This will short circuit the internal 
circuitry during a test.  
Example 
GROUND MODE = OFF, DUT floating 
High Voltage terminal
DUT
GPT-9600
stray 
resistance, 
capacitance
Return terminal
 
 
GROUND MODE = OFF, DUT grounded 
High Voltage terminal
DUT
GPT-9600
stray 
resistance, 
capacitance
Return terminal