PASS / FAIL Test Judgments for Auto Mode 
 
Background 
The test results for the Auto Mode testing are 
identical as the results from a single test, 
however:  
 
If one of the tests fails, then the FAIL 
judgment for that test will be shown on the 
display, even if the other test is judged as 
PASS. 
 
Only one test result can be viewed at any one 
time.  
 Note 
A test will be judged PASS when: 
 
The H and L limits (high and low limits) 
have not been tripped during the test time. 
A test will be judged FAIL when: 
 
Either the H or L limit has been tripped 
during the test time. 
 
The ARC current limit has been tripped 
during a test (if enabled).  
View Test 
Results~PASS  
1.  When both tests are judged as 
PASS, PASS will be displayed for 
the last test, the buzzer will sound 
once and the PASS indicator will 
be lit green.  
PASS
 
 
 
     
 
 
 
H = 2 2 . 0 m A  
 
 
C
 
m A
L
0
=
0 0
.
   
6 0 S  
 
 
 
 
 
= 0
T
       
   
CW
D
A CW
R  
I
-
R A
I
T M E
I
 
0 90
 
k V
S
A
P
 
S
A
m
 
1 00
 
 
 
 
2.  Turn the scroll wheel left to view 
the first test result. Turn the scroll 
wheel back to the right to view the 
last test result again.