GPT-9600 Specifications 
The specifications apply when the GPT-9600 is powered on for at 
least 30 minutes at 15˚C~35˚C.  
Specifications 
 
 
Environment 
Range 
Temperature 
Humidity 
Warranty  
15˚C ~ 35˚C 
≤70% (No condensation) 
Operation 
0˚C ~ 40˚C 
≤70% (No condensation) 
Storage 
-10˚C ~ 70˚C 
≤85% (No condensation) 
Installation Location 
Indoors at an amplitude of up to 2000m. 
 
AC Withstanding Voltage 
Output Voltage Range 
0.10kV~ 5.00kV  
Output Voltage Resolution  
10V 
Maximum Rated Load 
100 VA (5kV/20mA) 
Maximum Rated Current 
20mA  
0.01mA ~ 5mA(0.1kV≤V≤0.5kV) 
0.01mA ~ 20mA(0.5kV<V≤5kV) 
Output Voltage Waveform 
Sine wave 
Frequency 
50 Hz / 60 Hz 
Voltage Regulation 
± 1.5%+2 counts [Maximum rated load → no load] 
Output Voltage Accuracy 
± (1.5% of setting +2 counts)with no load 
Voltmeter Accuracy 
± (1.5% of reading+ 2 counts) 
Current Measurement Range  0.01mA~20.0mA  
Current Best Resolution 
10uA 
0.01mA(0.01mA~9.99mA) 
0.1mA(10.0mA~20.0mA)  
Current Measurement 
Accuracy 
± 2% of reading + 5 counts when I <1.00mA  
± 2% of reading + 3 counts when I ≥1.00mA 
Current Judgment Accuracy 
± 3% of setting + 5 counts when I <1.00mA  
± 3% of setting + 3 counts when I ≥1.00mA  
Window Comparator Method  Yes 
ARC DETECT 
Yes 
RAMP (Ramp Time) 
0.1S fixed 
TIMER (Test Time) 
OFF, 1S~180S 
GND 
ON 
Judgment Valid Range 
H/L Setting:  0.10mA to 9.99mA 
                      10.0mA to 20.1mA 
Continued…