Gw Instek GPT-9602 Insulation measuring device, GPT-9602 Manuel D’Utilisation

Codes de produits
GPT-9602
Page de 21
 
 
GPT-9600 Series Quick Start Guide 
14
 
Go to EDIT Status 
 
Press the EDIT key. 
EDIT
 
 
Set the Test Voltage 
 
When in the READY mode, turn the scroll wheel 
knob to set the test voltage.  
 
Start Testing 
 
Press the START button to begin testing. 
 
The tester will go into TEST status. 
START
 
 WARNING 
If Double Action is active, the START button must 
be pressed within 500ms after the STOP button 
was pressed to be able to start testing. 
 WARNING 
If INTERLOCK is set to ON and the interlock key is 
not connected to the SIGNAL I/O port, 
INTERLOCK OPEN will be displayed on the screen, 
preventing the test from starting. See page 19 for 
the Common Utility settings. 
EDIT Status 
 
Description 
EDIT status is used to edit the test parameters 
(excluding the test voltage). Pressing the SAVE 
key will save any changes and return to the 
READY status. Pressing the ESC key will cancel 
any changes and return to the READY status. 
Pressing the UTILITY key will enter the TEST 
Utility settings.