Pre-Tests and Protection 
 
Function Tests, 
protections 
Notes
Switching 
ON the 
instrument 
-
 Grounding of mains 
plug PE terminal 
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE 
DISCONNECTED! TURN OFF NOW!” message will be 
displayed and any further operation will be disabled. 
Switch off the tester immediately, danger! 
-
 Fuse F2
Pre-test.
If the F2 fuse is blown, then “FUSE 2!” message will be 
displayed and any further operation will be disabled. 
R
PE
 
-
 U
EXT PE/PROBE1
 
3.75 V  Pre-test.
If an external voltage higher than 3.75 V is detected 
before starting the test, then “EXTERNAL VOLTAGE!” 
message will be displayed and further R
PE
 tests will be 
disabled.  
-
 I
SUB INT
 
0.15 mA 
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g. 
due to blown internal fuse, then “ERROR 1” message 
will be displayed and further tests will be disabled. 
-
 Fuse F3
Pre-test and during the test.
If the F3 fuse is blown, then “FUSE 3!” message will be 
displayed and further R
PE
 tests will be disabled. 
-
 U
EXT PROBE1/GND 
50 V 
-
 U
EXT PROBE2/GND 
50 V 
Pre-test and during the test.
If an external voltage higher than 50 V is detected 
before starting the R
PE
 test or during it, then “EXTERNAL 
VOLTAGE!” message will be displayed and further R
PE
 
tests will be disabled. 
-
 U
EXT PE/PROBE1
 
7.60 V  During the test.
If an external voltage higher than 7.6 V is detected 
during the R
PE 
test, then the test will be stopped and 
“EXTERNAL VOLTAGE!” will be displayed. 
R
ISO
 
-
 U
LN/PE
 
50 V 
Pre-test.
If an external voltage higher than 50 V is detected 
before starting the R
ISO
 test, then “EXTERNAL 
VOLTAGE” message will be displayed and further R
ISO
 
tests will be disabled. 
-
 I
SUB INT
 
0.15 mA 
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g. 
due to blown internal fuse, then “ERROR 1” message 
will be displayed and further tests will be disabled. 
-
 U
EXT PROBE1/GND 
50 V 
-
 U
EXT PROBE2/GND 
50 V 
-
 U
EXT PE/GND 
50 V 
-
 U
EXT LN/GND 
50 V 
Pre-test.
If an external voltage higher than 50 V is detected 
before starting the R
ISO
 test, then “EXTERNAL 
VOLTAGE” message will be displayed and further R
ISO
 
tests will be disabled. 
I
SUB
 
-
 I
SUB INT
 
0.15 mA 
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g. 
due to blown internal fuse, then “ERROR 1” message 
will be displayed and further tests will be disabled. 
-
 U
EXT PROBE1/GND 
50 V 
-
 U
EXT PROBE2/GND 
50 V 
-
 U
EXT PE/GND 
50 V 
Pre-test.
If an external voltage higher than 50 V is detected 
before starting the I
SUB
 test, then “EXTERNAL VOLTAGE” 
message will be displayed and further I
SUB
 tests will be 
disabled.