Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 Fiche De Données

Codes de produits
LCR-816
Page de 91
  
BASIC 
MEASUREMENT 
37 
  
1.1kHz 
 
 
 
The nearest frequency will be selected from the 
504(LCR-281) nominal frequencies, and 
updated in the display. Here, the nearest 
frequency to 1.1kHz is 1.0909kHz. 
 
 
 
Note 
After the test frequency has been changed, the zeroing 
must be performed again. See page 24 
Set measurement voltage 
 
Background 
Along with frequency, voltage can be set. Make 
sure the appropriate voltage is selected, according 
to the component characteristics. 
 
Range 
5mV ~ 1.275V (5mV steps) <200kHz 
 
 
100mV ~ 1.275 (5mV steps) @200kHz 
Voltage setting 
1.  From the main menu, press MENU (F5) 
followed by SETTING (F3) and VOLT (F2) 
  
 
2.  Enter the voltage using the numerical keys, and 
then press ENTER. 
  
100mV 
 
   LCR-800 
User 
Manual 
38
 
  
1V 
 
 
 
 
 
 
The voltage is updated in the display. If the 
voltage entered is outside the allowable voltage 
range, the nearest voltage is selected. 
 
3.  Press (F5) EXIT to exit the Setting 
menu. 
 
Set PPM for D/Q measurements 
 
Background 
Dissipation and Quality Factor (D/Q) 
measurements can be shown in parts per million 
(PPM) if D/Q is less than 0.0100. This increases the 
resolution by a factor of 100. The units of D and Q 
are dimensionless and are expressed as a decimal 
ratio with a multiplier of 1,000,000.  
 
Ensure the operating mode has a D or Q component. 
See page 33. 
Panel operation 
1.  Press 4/PPM to turn PPM on or off 
for all D/Q measurements