145 
Replacing the AC Source Fuse 
 
Fuse Ratings 
Type 
Rating 
0.125AT 
100VAC, 120VAC 
0.063AT 
220VAC, 240VAC 
 Note 
Only replace the fuse with a fuse of the correct type 
and rating.  
Steps 
1.  Turn the DMM off and take out the power cord. 
2.  Remove the fuse socket using a flathead 
screwdriver. 
 
3.  Remove the fuse in the holder and replace with 
the correct type and rating. 
 
 
4.  Ensure the correct line voltage is lined up with 
the arrow on the fuse holder. Insert the fuse 
socket.