Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Fiche De Données
Codes de produits
BK-8500
0x40 Set one of the step's current and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the current in units of 0.1 mA
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
0x41 Read one of the step's current and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the current in units of 0.1 mA
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
0x42 Set one of the step's voltage and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the voltage in units of 1 mV
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
0x43 Read one of the step's voltage and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the voltage in units of 1 mV
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
0x44 Set one of the step's power and time values
Byte
offset
Meaning
3 to 4
2 byte little-endian integer specifying which step number in the list
5 to 8
4 byte little-endian integer specifying the power in units of 1 mW
9 to 10
2 byte little-endian integer specifying the step timing in units of 0.1 ms
11-24
Reserved
8500 DC Load Series
Version: September 2, 2009