Kern Precision scales PCB 6000-1 Weight range 6 kg Readability 0.1 g mains-powered, rechargeable Silver PCB 6000-1 Fiche De Données

Codes de produits
PCB 6000-1
Page de 37
PCB-BA-e-1315 
33 
 
Baud rate 
 
 
 
 
 
The baud rate defines the transfer speed vie the interface, 1 Baud = 
1 Bit/second. 
 
  In weighing mode keep the PRINT key pressed until [Unit] 
appears. 
  Press the MODE 
key several times until „bAUd“ is displayed. 
  Acknowledge using SET key, the current setting is displayed. 
  Use MODE key select the desired settings 
 
9600  4800  2400  1200  19200 
  Use the SET key to confirm selection. The balance returns to 
weighing mode. 
 
 
10  Data output RS 232 C 
10.1  Technical Data 
 
 
8-bit ASCII Code 
 
1 start bit, 8 data bits, 1 stop bit, no parity bit 
 
Baud rate selectable at 1200, 2400, 4800 , 9600 and 19200 baud  
 
Miniature plug-in necessary (9 pole D-Sub) 
 
For  operation  with  interface  faultless  operation  is  only  ensured  with  the  correct 
KERN 
– interface cable (max. 2m) 
10.2  Pin allocation of balance output bushing: 
 
Front view:
 
 
 
 
      1
 
 
                                    
9             6    
 
 
Pin 2: 
Transmit data 
Pin 3: 
Receive data 
Pin 5: 
Signal ground