5 Funct
ion
s
14
5.1
Measurement input
Symbol
Notes
C111
Transducer/probe (measurement input)    
C112
Unit of measurement (°C/°F)/decimal places in display    
SCL
Start/end value of value range 
for standard signals and resistance transmitter    
Example: 0—20 mA
20— 200°C: 
SCL
 = 20 / 
SCH
= 200
SCH
OFFS
Correction of the actual value    
Actual-value correction permits adjusting a measured value upwards or downwards by a
programmable value (offset). For a 2-wire circuit, lead compensation can thus be implemented 
in the software (for Pt100 probes, for example, a lead resistance of 0.39
⍀ corresponds to a 
temperature deviation of 1°C).
Examples:
measured value 
offset 
displayed value
294.7 + 
0.3 
295.0 
295.3 
 - 0.3 
295.0 
dF
Filter time constant (damping), for adapting the digital input filter (0sec = filter off).
The value range of 
dF
 is 0.0 — 100.0sec; factory setting: 0.6sec.
if 
dF
 is large: 
- high damping of interference signals
- slow reaction of the actual-value display to changes in the actual value
- low limit frequency (2nd order low-pass filter)