Technical Data
② according to IEC-1010; UL 3101
Type
Nominal voltage (±10 %)
Nominal frequency
Power draw
Max. start-up current for 5 sec.
Protective class
Circuit breaker:
(MDA made by Schurter)
Pollution degree
Overvoltage installation category
Heat emission 
(max.)
Approval
-1
-2
-3
100 V AC
120 V AC
230 V AC
50/60 Hz
60 Hz
50 Hz
1900 VA
1900 VA
1900 VA
35 A eff.
35 A eff.
25 A eff.
I
I
I
15A M3
15A M3
15A M3
2
2
2
II
II
II
1900 J/s
1900  J/s
1900 
 
J/s
CE/c_CSA_us
CE/c_CSA_us
CE/c_CSA_us
All specifications related to temperature are valid only up to an ambient temperature of 18 °C
to 35 °C and a relative humidity of no more than 60%.
Microtome
Type:
Rotary microtome,
encapsulated
Section thickness range
1 to 100 μm
Trimming range:
Clinic
10 - 40 μm 
Research
1 - 600 μm 
Total specimen feed
25 mm + 1 mm
Vertical stroke
59 mm ±0.5 mm
Specimen retraction
20μm (can be deactivated)
Maximum specimen size
50 x 80 mm
Cutting speed
Slow: 0-50 strokes/min
Fast: 0-85 strokes/min
Vmax: 85-90 strokes/min
Specimen orientation
8° (x, y-axis)
Electric coarse feed
Slow
300 μm/s
Fast
900 μm/s
Cryostat
Dimensions:
Width (w/o handwheel)
700 mm
Width (with handwheel)
835 mm
Depth (cabinet only)
850 mm
Height (total)
1215 mm
Working height (armrest)
1025 mm
Weight:
Weight (w/motor and suct.)
193 kg
Weight (w/motor, w/o suct.)
185 kg
Weight (w/o motor, w/ suct.)
183 kg
Weight (w/o motor, w/o suct.)
175 kg
Weight (w/ specimen head cool.)
165 kg
Weight (w/o specimen head cool.)
145 kg
For "Site Requirements," refer to item 6.1
(page 20).
Lamp
50/60 Hz-Version:
Osram DULUX L 18 W/840
General information
Operating temperature range: 18 °C to 35 °C
Temperature range
during storage:
+5 °C to +55 °C
Relative humidity:
max. 60 %,
non-condensing
Storage humidity:
< 60 %