Leica CM1950 Manuel D’Utilisation

Page de 74
33
Leica CM1950  – Cryostat
7.
Instrument Controls
Coarse feed functions
The two-speed electric coarse feed is designed to quickly position the spec-
imen in relation to the knife. The buttons with double arrows move the spec-
imen at 900 μm/s; the buttons with single arrows move it at 300 μm/s.
Retracting the specimen head from the knife
Pressing once starts the rapid backwards movement to the rear end
position (Home position).
The LED (1) flashes while the specimen head is in motion.
The LED (1) lights up when the rear end position (Hp.) has been reached.
The retraction can be stopped by pressing one of the coarse drive buttons.
The rapid backwards movement to the rear end position (Hp.) starts.
The motion continues as long as the button is held.
A brief press of the button retracts the specimen by 20 μm.
Manual sectioning mode
Select ROCK operating mode (LED active)  – retraction must be enabled!
For sectioning, turn the handwheel a short distance (approx. 1/4 turn)
forwards and back (rocking mode) – only possible at rear (handwheel in
approx. 12 - 3 o'clock position).
Every change in rotation direction is electronically detected and automati-
cally translated into a specimen feed or retraction movement.
1
2
slow
fast
slow
fast
Advancing the specimen toward the knife
Start the slow forwards movement to the knife.
To feed the specimen, press and hold the button.
Pressing the button briefly results in a feed motion of 20 μm.
Start the fast forwards movement to the knife.
The LED (2) flashes while the specimen head is in motion.
The LED (2) lights up when the forward end position has been reached.