Leica CM1950 Manuel D’Utilisation

Page de 74
45
Leica CM1950 Cryostat
8.
Working with the Instrument
Activate trimming mode.
Select the trimming thickness.
Switch on the VAC and select a low extraction level (between 1 and 2).
Start trimming by manually moving the handwheel or press the
RUN/STOP and ENABLE buttons  simultaneously to start motorized
sectioning.
Use a brush to apply the section to a precooled specimen slide, then
warm it from below using your finger.
Move the knife guard forward after removing the section.
Remove the knife or disposable blade (using the blade ejector!).
Insert the knife into the knife case.
8.4.3 Cutting with Extraction – Anti-roll Guide installed
Shut off VAC (LED in VAC button goes out).
Switch from trimming to sectioning mode (important for section stretch-
ing, as the valves work differently than in trimming mode.)
Set the desired section thickness.
Switch on the VAC and begin at level 1. If the section does not stretch
correctly, turn the VAC knob higher in small increments.
Once the desired section is on the pressure plate, switch off the VAC!
Carefully fold the anti-roll guide to the side and remove the section from
the side.
After removing the section, wipe off moisture/condensate from the
pressure plate – otherwise, the next sections will become
jammed.
Cutting with extraction without anti-roll guide (brush technique) is
not possible, as the position of the pressure plate means that no
suitable air flow is obtained.