Cardinal Gates VICTOREEN 948A-50 Manuel D’Utilisation

Page de 16
 
4-1
Section 4 
Maintenance, Calibration,  
and Troubleshooting 
 
4.1 Maintenance 
The 948A-40 Remote Alarm and 948A-50 Remote Alarm/Meter require no routine maintenance other than periodic 
calibration for the 948A-50. Refer to calibration for further information. 
 
4.2 Calibration 
The 948A-50 requires calibration before it is initially placed into service. In addition, it should be recalibrated at regular 
intervals during routine service. The length of time between calibration intervals should be determined by the 
customer. Use the following procedure to perform the calibration: 
1.  Using radios, field telephones, or other suitable communication devices, establish communications between a 
technician at the ratemeter location and another at the remote alarm/meter location. 
2.  To gain access to the meter adjustment potentiometer, open the front panel of the remote alarm/meter. The 
potentiometer is mounted on the rear of the panel meter. 
3.  Adjust the potentiometer until the reading shown on the panel meter agrees with the reading shown on the 
ratemeter within 
±5%. 
4.  This concludes the calibration procedure for the 948A-50 Remote Alarm/Meter. 
 
4.3 Troubleshooting 
 
 
Extreme care must be used when troubleshooting a 
system that has power applied. All standard 
troubleshooting precautions apply. 
 
 
 
Once a problem has been located, remove all power 
before continuing with the repair. 
 
 
 
Personnel performing the following procedure must be 
familiar with the operation of the system and the 
location of each piece of equipment used. 
 
WARNING
WARNING
CAUTION