SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B Manuel D’Utilisation

Page de 48
 
 7-13 
10. If storing the pump for more than 12 hours fill each pump head with 
100% isopropanol then seal the inlets and outlet with the supplied 
plugs. 
5.7 Lubrication 
The Binary pump has modest lubrication requirements. The 
bearings in the pump housing and piston carrier are permanently 
lubricated and require no maintenance. A small dab of light grease 
such as Lubriplate 630-AA on the cam is the only recommended 
lubrication. Be sure not to get lubricant on the body of the piston 
carrier, as this can retard its movement and interfere with proper 
pumping. 
NOTE:  Keeping the interior of the pump free of dirt and dust 
will extend the pump’s useful life. 
 
5.8 
Other Pump Maintenance 
 
  
The internal components of the pump can be accessed by 
removing the pump drawer.  To do so, turn off power at front switch.  
Unplug power cord.  Ensure there is no power to the system.   
Disconnect the inlet line and disconnect pump outlet at the “T” 
Connector Block above.  Unscrew the two thumbscrews on the upper 
portion of the drawer until fully released (counterclockwise).  Pull the 
drawer out as shown on next page. 
 
Note:  Drawer can be removed completely.