SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B Manuel D’Utilisation

Page de 48
 
 A-1 
APPENDIX  A 
A.1 
Rear Panel Serial Communications Port - Pump 
 
An RS-232C modular jack is provided on the back panel.  A 
computer, with appropriate software, can be used as a remote 
controlling device for pump operation via this connection. 
A.1.1 Hardware Implementation 
The REMOTE INPUT serial communications port is configured 
for 9600 baud, 8 data bits, 1 stop bit, and no parity.  The connector is a 
standard RJ-11 modular telephone type jack.  When looking at the 
connector on the rear panel of the pump, pin 1 is at the top and pin 6 is 
at the bottom.  The pin-out is: 
 
 
Pin 
Function
 1, 
Ground 
 
DSR (Handshaking input to pump) 
 
RXD (Serial data input to pump) 
 
TXD (Serial data output from pump) 
 
DTR (Handshaking output from pump) 
 
Special wiring considerations:  Use the following chart for 
interfacing the pump's serial communications port to either a 25-pin or 
a 9-pin COM port on an IBM-PC type computer. 
 
 
Pump (RJ11) 
Signal 
 
IBM (DB25)
a
 
IBM (DB9)
b
 
 1, 
Ground 
 
 2 
DSR  
20 
 3 
RXD 
 
 
 4 
TXD 
 
 5 
DTR  
 
 
a
 Jumper pins 4, 5, and 8 on DB25. 
 
b
 Jumper pins 1, 7, and 8 on DB9. 
 
 
Part Description 
Part Number 
 
 Modular 
Cable 
12-0677 
 
Adapter RJ-11 to DB9 
12-0672 
 
Adapter RJ-11 to DB-25 
12-0671 
 
A.1.2 Hand-Shaking 
The pump uses hardware handshaking.  The pump will not 
transmit on the TXD output if the DSR input is at a low logic level.  
And, the pump will not receive on the RXD input when the DTR