SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B Manuel D’Utilisation

Page de 48
 
 A-5 
the EVENT 1 terminal and the EVENT 2 and EVENT 3 terminals is 
affected.  EVENT 2 is Normally Closed (connected to EVENT 1) until 
a fault occurs and then opens.  EVENT 3 is Normally Open (not 
connected to EVENT 1) until a fault occurs and then closes. 
 
A.2.1.1  Upper and Lower Pressure Limit Range 
 (pump A  only)
 
 
The pressure sensing transducer provides accurate, wide range 
pressure monitoring.  Because of the sensitivity of the transducer, 
the zero reading may shift up to 0.1% of the full pressure scale 
over years of operational use.  The user should also be aware that 
the resistance to flow  of the fluid being pumped through the 
tubing and fittings may cause the pressure to variy with the flow 
rate and the viscosity of the mobile phase employed.  
 
If absolute accuracy is needed for the pressure safety limits: 
 
 
1.  Disconnect the column from pumping system and operate the 
pump with the mobile phase and flow rate to be used in the 
analysis.  Observe the resulting pressure displayed on the pump 
readout.  The column will cause a pressure reading that adds to this 
basic reading due to system flow resistance.   
 
 
2.  Set the upper limit shut-off to a pressure equal to the basic 
reading plus the safe operating pressure for the column to be used.  
For example, if the basic pressure reading (without the column) is 
7 PSI and the safe limit for the column is 25 PSI, set the maximum 
pressure limit to 32 PSI or less. 
 
 
3.  If the mobile phase or flow rate is changed, reset the pressure 
limit as appropriate. 
 
 
4.  Note that a lower pressure limit is available to prevent 
continued operation in the event of a leak.  For proper operation, 
this must be set to a pressure higher than the basic pressure or it 
may not sense the reduced pressure. 
 
 
 
A.2.2  General Information on Inputs 
The pump's inputs are on the 10-pin terminal board connector. The 
pinout is: