VXi Computer Hardware VX1410A Manuel D’Utilisation

Page de 360
STATus Subsystem
3–96
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:QUEStionable:VOLTage:PTR
STATus:QUEStionable:VOLTage:PTR?
Sets or queries the power supply positive transition filter.
You can only set or clear bits 3 and 4 of the transition filter. Setting a bit enables
the corresponding power supply condition to propagate to the event system. Any
failures will display on the front panel readout. If you disable the bits, any failure
in the corresponding power supply will not be reported to the front panel readout
or to the event system.
Bits 0 – 2 and bits 5 – 8 of the transition filter are always enabled. Bit 15 always
returns a 0. The factory default setting is 32743. The +5 V Standby and +5 V
External are not monitored.
STATus:QUEStionable:VOLTage:PTR <Mask>
STATus:QUEStionable:VOLTage:PTR?
<Mask>
Query response
{<NRf> | <Nondecimal Numeric>}
 N 
 #HFFFF
<NR1>
Voltage positive transition filter bits
Definition
D3
+5 V Standby
D4
+5 V External
0
–222, “Data Out of Range”
You attempted to set the transition filter to an illegal value.
None
Command:
STAT:QUES:VOLT:PTR #H7FF7
Query:
STAT:QUES:VOLT:PTR?
Response:
32759
STATus:QUEStionable:VOLTage:CONDition?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands