Siemens Sitrans F FX300 Manuel D’Utilisation
6
OPERATION
56
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02 A5E02100423
3.3.3
HART TV
→
HART tertiary variable
Temp. ↑...
Pressure ↑...
Density ↑...^
3.3.4
HART 4V
→
HART quaternary variable
Temp. ↑...
Pressure ↑...
Density ↑...^
3.4
Set fluid and medium
3.4.1
Fluid 1
→
Fluid type
Steam ^
Steam
3.4.2
Medium
→
Process medium
Sat.Steam ↑...
Saturated steam
Sup.Steam ↑...^
Overheated steam
3.4.5
Dry. Fact.
→
(available if Medium =
Sat.Steam, see menu item
3.4.2)
Enter portion of dryness in saturated steam
0000001.00
0.85 TO 1 ↑…↑...
↑...^
Factor = 1 - %water mass
1 Density check
3.5
Pressure, temperature and density
3.5.1
T-sensor
→
Internal temperature sensor
No ↑...
No temperature sensor available
Yes ↑...^
Temperature sensor available
3.5.2
P-sensor
→
Pressure sensor
Intern ↑...
Internal pressure sensor
--- ↑...^
No pressure sensor available
3.5.3
Sat.P/T
→
(available if Medium =
Sat.Steam, see menu item
3.4.2)
Density calculation for saturated steam
Sat. Temp. ↑...
Density calculation for saturated steam from saturated
temperature
Sat. Pres. ↑...^
Density calculation for saturated steam from saturated
pressure
3.5.4
Temp. Opr.
→
(for saturated steam with
Sat. Temp. or Sup.Steam,
see menu items 3.4.2 and
3.5.3)
Operating temperature
°C
Unit ↑...^
Set temperature unit
0000000.0
°C ^
Operating temperature
Disp. on /
Disp. off ↑...^
Display temperature /
Do not display temperature
3.5.5
Pres. Opr.
(available for Sat.Steam
with Sat. P or Sup.Steam,
see menu items 3.4.2 and
3.5.3)
Operating pressure
Pa
Unit ↑...^
Set pressure unit
0000000.0
Pa ^
Operating pressure
Disp. on /
Disp. off ↑...^
Display pressure /
Do not display pressure
Level
Designation
Selection / entry
Explanation
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 56 Monday, June 22, 2009 8:17 AM