Siemens Sitrans F FX300 Manuel D’Utilisation

Page de 80
6
 OPERATION 
56 
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02   A5E02100423
3.3.3
HART TV
HART tertiary variable
Temp. ↑...
Pressure ↑...
Density      ↑...^
3.3.4
HART 4V
HART quaternary variable
Temp. ↑...
Pressure ↑...
Density      ↑...^
3.4
Set fluid and medium
3.4.1
Fluid 1
Fluid type
Steam ^
Steam
3.4.2
Medium
Process medium
Sat.Steam ↑...
Saturated steam
Sup.Steam ↑...^
Overheated steam
3.4.5
Dry. Fact.
(available if Medium = 
Sat.Steam, see menu item 
3.4.2)
Enter portion of dryness in saturated steam
0000001.00
0.85 TO 1     ↑…↑...
↑...^
Factor = 1 - %water mass
1 Density check
3.5
Pressure, temperature and density
3.5.1
T-sensor
Internal temperature sensor
No ↑...
No temperature sensor available
Yes      ↑...^
Temperature sensor available
3.5.2
P-sensor
Pressure sensor
Intern ↑...
Internal pressure sensor
---      ↑...^
No pressure sensor available
3.5.3
Sat.P/T
(available if Medium = 
Sat.Steam, see menu item 
3.4.2)
Density calculation for saturated steam
Sat. Temp. ↑...
Density calculation for saturated steam from saturated 
temperature
Sat. Pres. ↑...^
Density calculation for saturated steam from saturated 
pressure
3.5.4
Temp. Opr.
(for saturated steam with 
Sat. Temp. or Sup.Steam, 
see menu items 3.4.2 and 
3.5.3)
Operating temperature
°C
Unit  ↑...^
Set temperature unit
0000000.0
°C  ^
Operating temperature
Disp. on /
Disp. off ↑...^
Display temperature /
Do not display temperature
3.5.5
Pres. Opr. 
(available for Sat.Steam 
with Sat. P or Sup.Steam, 
see menu items 3.4.2 and 
3.5.3)
Operating pressure
Pa
Unit  ↑...^
Set pressure unit
0000000.0
Pa  ^
Operating pressure
Disp. on /
Disp. off ↑...^
Display pressure /
Do not display pressure
Level
Designation
Selection / entry
Explanation
MA_Siemens_Sitrans_FX300_R02_en_PRT.book  Page 56  Monday, June 22, 2009  8:17 AM