Thank you for purchasing the Temple Touch Thermometer.  
Temple Touch enables accurate temperature measurements from the temple area in approximately 6-8 seconds, 
using simple steps.
Table of contents:
Safety warnings
Section 1: 
Quick Start
Section 2: 
measuring guidelines
Section 3: 
Directions for use 
Section 4: 
Installing and replacing the battery
Section 5: 
About body temperature measurement
Section 6: 
Troubleshooting and special indications
Section 7: 
Care and cleaning 
Section 8: 
Technical Information
Section 9: 
Warranty and Service Information 
IMPORTANT!  Read user manual before using the thermometer.
Safety warnings:
1.  Keep the thermometer out of the reach of children 
2.   Do not drop the thermometer or expose to heavy shock. The unit is not shock proof.
3.   Do not immerse thermometer in water or any other liquid. The unit is not water proof.
4.   Do not store the unit in direct sunlight, at a high temperature, in high humidity or dust.  
Performance of the unit may be  degraded. 
5.   Stop using the thermometer if it operates erratically or if display malfunctions.
6.  Clean the thermometer probe before storing (see section 7)
7.  Do not attempt to disassemble the unit except to replace the batteries.
1. QUICK START
1. Press and release the POWER button. 
2. Unit will beep twice when ready. 
3. Hold firmly to temple, wait for next beep. 
4. Read results.
2. Measuring guidelines:
1.   Measurement body location is the temple, in proximity to the temporal artery, between the eye socket and the 
hairline.
2.   Do not place the thermometer on scarred tissue compromised by skin disorder, open sores or abrasions and 
patients in trauma.
3.   Consult your physician when using drug therapies that may raise the local skin temperature in the forehead; 
this may lead to incorrect reading. 
4.   Do not place the thermometer on skin exposed to direct sunlight, fireplace heat, air conditioner flow and 
cold/warm compress therapies.
5.   
Wait  at  least  two  minute  between  consecutive  measurements  for  the  unit  to  readjust  to  room 
temperature.
6.   If  the  thermometer  has  been  stored  at  a  cold  temperature,  allow  the  unit  to  warm  naturally  to  room 
temperature before using.
7.   Performance of the device may be degraded if operated or stored outside stated temperature and humidity ranges. 
3. Directions for Use:
Before measuring:
a.   Use this thermometer on a clean dry forehead. Wipe away any perspiration prior to use.
b.   Move hair aside from the temporal area and apply the probe to bare skin.
Taking Temperature:
1.   Press and release the ON/OFF 
 button, you will hear one beep.  
All segments will be displayed. 
2.   The previous result will be displayed for 2 seconds. 
3.   Wait for the hourglass icon to disappear, then two beeps will be heard  
to indicate that the thermometer is now ready to measure
 
Do not try to take a measurement before the two beeps are heard.
4.   Place the thermometers probe in the center of the temple area between the  
eye socket and the hairline. Gently press and hold it against the temple to  
ensure good contact with the skin is maintained. 
   
Correct placing and good contact with the skin ensure reliable readouts.
5.   As soon as the thermometer is in place, the measurement will start automatically  
A running circle will appear indicating that the measurement is progressing  
successfully. 
Make sure that the thermometer and the patient do not move  
while measurement is in progress
6.   A temperature reading will be displayed within 6-8 seconds. One long beep will be heard  to indicate that 
measurement is complete and temperature will be displayed for about 1 minute
7.   Press and release the On/Off  button to turn the thermometer off, or it will shut off automatically within 1 
minute.
   If you need to take another reading – wait 2 minutes after turning the thermometer off and then begin 
again. 
Setting the operating scale °F or °C
The device setting is by default Fahrenheit (°F).
To change the scale to °C, or back to °F:
1.   Press and hold the ON/OFF   button for 5 seconds – on the screen you  
will see the hourglass icon blinking and the last measured temperature.
2.   Once the last reading showed on the screen disappears, the scale  
icon (
0
F, 
0
C) changes as well.
3.   Release the button, the hourglass icon will disappear in 5 seconds and  
2 beep sounds are heard, indicating that the device is ready for use. 
4. Installing and replacing the Battery:
The thermometer uses 2 standard AAA 1.5v batteries. To install/replace:
1.  Remove the battery cover by sliding it out of its housing.
2.   Firmly attach the batteries to their place, make sure polarity is correct. (see engraving on the bottom of 
battery compartment).
3.   Slide the battery cover back into place.
   IMPORTANT: It is recommended to remove the batteries from the thermometer when it is not used for a long 
period.
   WARNING: Do not dispose of batteries in fire, recharge, put in backwards, disassemble, mix with used or 
other battery types – they may explode or leak and cause injury. Waste batteries should be deposit in a 
batteries recycling trash only. 
5. About body temperature measurement:
How does the thermometer work?
The accurate measurement is obtained by using a unique probe, which 
 measures the heat flow from the blood vessels to the  
skin and converts it to body temperature.
Information about Body Temperature
What is “normal” body temperature?
“Normal” body temperature varies from person to person, changes according to time of day and body location. 
Hot or cold baths, physical exercise, smoking and drinking, hot or cold drinks may change body temperature 
by one degree or more.
 
 
 
The following table shows the temperature ranges:
What is fever?
Average normal temperature ranges
Forehead
Over 37.8°C/100.0°F
36.1°C–37.8°C / 97.0°F-100.0°F
Rectal*
Over 38°C/100.4°F
36.3°C–38°C /  97.3°F-100.4°F
Oral (mouth)*
Over 37.8°C/100.0°F
36.1°C–37.8°C /  97.0°F-100.0°F
Axilla (armpit)*
Over 37.2°C/99.0°F
35.2°C–37.2°C /  95.4°F-99.0°F 
* Fever definitions according to the American Medical Association
The thermometer is designed to measure the temperature in the forehead, in proximity to the temple area, 
between  the  eye  socket  and  the  hairline.  It  is  important  to  know  that  the  thermal  characteristics  of  each 
measurement  body  location  –  temporal,  axillar,  oral  or  rectal  -  are  unique.  The  variance  in  temperature 
measurement in different body locations is due to the human body and not the thermometer.
IMPORTANT:  It  is  recommended  to  measure  the  body  temperature  when  healthy  to  establish  the  specific 
“normal” temperature
6. Troubleshooting and special indications
The  thermometer  has  been  designed  to  display  feedback-warnings  to  ensure  proper  usage  and 
accurate  temperature  measurement.  When  an  error  appears,  the  device  will  "beep"  5  times,  
and then the thermometer will turn off automatically. In the event of a malfunction or incorrect temperature 
measurement, the following warnings will be displayed:
Indication:    The measured temperature is below 95°F (35°C) and the display presents: LO. 
Instruction:  1. Turn the unit OFF and then ON again. 
 
2.  Retake temperature using proper placement technique (see section 3).
Indication
 The measured temperature is above 107.6°F (42°C) and the display presents: HI
Instruction:   Turn the unit OFF and then ON again. To verify the reading, retake temperature  
using Proper technique (see section 3).
WARNING! A high fever condition requires seeking immediate medical attention
Indication 1
:  The thermometer is too cold or too hot to operate. 
Instructions:   1. Verify room temperature within product range  
60.8°F-104°F (16° C - 40° C)  
 
 2. Retake temperature using proper technique (See section 3).
Indication 2:   Temperature did not register because of poor contact, patient  
movement  or poor placement.
Instruction:   Turn the unit OFF, wait 2 minutes, turn it ON again and retake  
temperature using proper technique.( See section 3).
Indication 3:  Unit malfunctioned.
Instruction:   Do not use the thermometer – contact service. (See section 9) 
Indication
 Low battery - Battery icon is on continuously.
Instruction:    The battery has enough remaining life for approximately 10 more  
measurements
Indication
 Drained battery – only Battery icon is on when thermometer is turned  
on and the device turns itself OFF.
Instruction:   The battery is drained - replace battery.(See section 4)
7. Care and Cleaning
* Wipe the tip and probe with ReliOn
®
 alcohol swab or a cotton swab moistened with 70% isopropyl alcohol 
before & after each use.
After  cleaning,  wait  at  least  2  minutes  before  taking  a  measurement  for  the  unit  to  readjust  to  room 
temperature.
* Use only a soft dry cloth to clean the body of the thermometer (if necessary). Never use an abrasive cleanser 
or submerge the thermometer in water or other liquids. 
* Do not expose the thermometer to extreme temperatures, humidity, direct sunlight or shock. Store at room 
temperature.
8.  Technical Specifications
Feature
Performance / Data
Measurement time
6-8 seconds
Measurement 
locations
Temple  area of the forehead
Measurement range
95°F to 107.6°F (35°C to 42°C)
Operation environment
60.8°F-104°F (16° C - 40° C) Humidity: 15-95% (non condensing)
Storage environment
-4°F to 122°F (-20° C - 50° C), Humidity: 15-95% (non condensing)
Measurement readout
Celsius or Fahrenheit
Accuracy
complies with ASTM E 1112-00 standard requirements
Display
TN LCD 0.63" X 0.44" (16 mm X 11 mm) 
Battery
2 AAA 1.5volt batteries
Battery life
Minimum 1000 full duty cycles over 24 month period 
Weight
2.1oz (60gr) including batteries
Dimensions
4.7" x 1.17" x 1" (12 cm x 3.0 cm x 2.4 cm ) 
Memory
Last reading display when turned on
Auto shut off
After 60 seconds
Case material
Medical grade ABS plastic
Probe material
Medical grade Stainless steel ,medical grade ABS plastic 
Warranty
1 year
9. Warranty and Service Information:
This  thermometer  has  a  one  year  warranty  against  manufacturers  defect  under  normal  use.  Following  the 
directions closely will ensure years of dependable operation. If the thermometer does not function properly, first 
check the batteries (see section 4). Replace if necessary.
If the thermometer does not function properly, please write to us at costumerservice@medisim-usa.com, call us 
at 1-888-723-3659 or visit us at www.relion.com. Our representatives will provide further instructions on how to 
correct the trouble or will ask you to return it for repair/replacement.
* Internally powered device
* Type - BF Applied part.
* Degree of protection against ingress of water: Ordinary
 
ReliOn
®
 is a registered Trademark used under license by Medisim Ltd. G.G Communication Center, Neve Iian 
90850 Israel.
Distributed by Wal-Mart Stores, Bentonville AR 72716.
Protected by US patent # 6,280,397 B1 and patent application # 11,444,710.
Temple Touch used with permission of Medisim, Ltd.
Made in China
  Read the Instructions for Use
  Applied part type BF equipment 
Instructions for Use  
    
 Temple Touch Thermometer
 
1
1
Skin
Energy  
Flow
Blood 
vessels
3
4
Power control
 
non-invasive  
sensor
2
5
2