Hitachi Microscope & Magnifier S-3400N Manuel D’Utilisation

Page de 466
3.3.2 
3 - 17 
(1)  For accelerating voltage, select the desired accelerating voltage level from the list displayed 
in the accelerating voltage Vacc input section.    If necessary, the Up/Down 
 buttons can 
be used to set the defined value in 100V increments.    If desired, a specific numerical value 
can be entered from the keyboard (use the Enter key to set the value). 
 
(2)  Set a filament current, Filament.  Select DETAIL on the Optics tab on the Control Panel, 
and press AFS with Mid selected. 
 
 
 
 
Fig. 3.3-6    Setting a Filament Current 
 
NOTICE:  The filament current, Filament, can be set automatically by pressing either the Auto 
Filament Saturation AFS function or the Auto Beam Setting ABS button. 
 
NOTICE:  Because the previous setting value is memorized in Filament, AFS need not be done 
every time.    Execute it when Filament setting changes.   
 
Also, the filament current Filament can be adjusted on a customized basis by moving the slider. 
Normally, Hitachi recommends the use of an automatically set filament current Filament by 
means of the Auto Filament Saturation AFS function or the Auto Beam Setting ABS option to 
Image a specimen. 
When the AFS function is running, the screen freezes, and a dialog message appears. When 
the screen changes to a RUN state, it is an indication that the auto adjustment has finished. 
 
NOTICE:  The Auto Filament Saturation AFS should be set to High for high magnification 
imaging, and Low to ensure long filament life.