Emerson Transmitter 350 Mode D’Emploi

Page de 35
Section
8
Transmitter Calibration and
LTM Troubleshooting
8.0  Calibration
There really is no need to re-calibrate this 
instrument.  The LTM-Series arrive calibrated 
to specified measuring ranges at order 
placement.
If the need arises and a recalibration is deemed 
necessary please follow these instructions carefully.
Level simulation will be required at 0% and 
100% positions.
1. Verify the “probe length” is not the same as 
the” measuring range” desired.  If so please 
change accordingly.
See Sections 2.1 and 7.1.
2. Go to the “Change Range” menu and set 
the desired URV (measuring range).
3. Go to the “Trim Sensor” menu and enter.
4. At the “Trim Zero” screen place the level (or 
magnet/float) at 0% (4mA point) and then 
press enter.
5. At the “Trim Span” screen place the level 
(or magnet/float) at 100% (20mA point) and 
then press enter.
6. “Exit” the configuration menus and the 
instrument should be calibrated.  If not 
begin to troubleshoot or consult factory for 
assistance.
Error Preventative Calibration: The LTM-250/350 
transmitters are protected from incorrect 
calibration so if something is done incorrectly it will 
display “Span Error” on the LCD for 5 seconds and 
then return to the beginning of the “Trim Sensor” 
menu.
Recommendation:  Again there is no need to 
recalibrate the instrument simply go to the 
“Change Range” menu and adjust the LRV or URV 
as desired.
8.1  Troubleshooting
Magtech manufactures custom built products so 
each transmitter has been specifically designed to 
fit on a magnetic level gage or be inserted into a 
tank/vessel and has been approved by the end user 
at some point during the ordering/purchasing 
process.  Each transmitter has been factory 
calibrated to meet end user requirements so there 
is no need to recalibrate these transmitters unless 
the requirements have changed from the time of 
purchase/delivery to installation.  If the need arises 
to calibrate these units to a new range please 
carefully read and follow the calibration procedure 
in the instruction and operation (I/O manual) or 
consult the factory.  Always feel free to contact the 
factory if any questions or uncertainties arise.  The 
following guide has been established to guide the 
end user through some troubleshooting procedures 
if the need arises.
There are 3 basic scenarios that can potentially 
cause issues.  They are:
1. Physical damage to the sensor probe.  
Accidents can happen during the installation 
process and if the sensor is permanently bent, 
bowed, or kinked it will no longer work 
properly.  Some symptoms of damage to the 
sensor probe are:
a. Erratic output, unit jumping to full scale 
(alarm mode) randomly, usually around 
the damage point.
b. Unit in constant alarm condition
c. Unit going into alarm condition after a 
certain point.
If the unit arrives damaged please contact 
factory immediately.  
28