FMC Corporation - Talstar 80 Manuel D’Utilisation
Troubleshooting
Proline Promass 80 PROFIBUS PA
74
380
S: UPPER FREQ. LIM
$: # 380
$: # 380
0x0F
BAD
Device Failure
Constant
Frequence limit 2
Cause:
The measuring tube oscillation frequency is outside
the permitted range.
The measuring tube oscillation frequency is outside
the permitted range.
Causes:
Measuring tube damaged
Sensor defective or damaged
Measuring tube damaged
Sensor defective or damaged
Remedy:
Contact your Endress+Hauser service organization.
Contact your Endress+Hauser service organization.
381
S: FLUIDTEMP.MIN.
$: # 381
$: # 381
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
Fluid Temperature
Min.
Min.
Cause:
The temperature sensor on the measuring tube is
likely defective.
The temperature sensor on the measuring tube is
likely defective.
Remedy:
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is
correctly plugged into the amplifier board
• Remote version: Check sensor and transmitter
terminal connections No. 9 and 10
→ Page 29 ff.
→ Page 29 ff.
382
S: FLUIDTEMP.MAX.
$: # 382
$: # 382
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
Fluid Temperature
Max.
Max.
383
S: CARR.TEMP.MIN
$: # 383
$: # 383
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
Carrier
Temperature Min.
Temperature Min.
Cause:
The temperature sensor on the carrier tube is likely
defective.
The temperature sensor on the carrier tube is likely
defective.
Remedy:
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is
correctly plugged into the amplifier board
• Remote version: Check sensor and transmitter
terminal connections No. 11 and 12
→ Page 29 ff.
→ Page 29 ff.
384
S: CARR.TEMP.MAX
$: # 384
$: # 384
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
Carrier
Temperature Max.
Temperature Max.
385
S: INL.SENS.DEF.
$: # 385
$: # 385
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
Inlet Sensor Defect
Cause:
One of the measuring tube sensor coils (inlet or
outlet) is likely defective.
One of the measuring tube sensor coils (inlet or
outlet) is likely defective.
Remedy:
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is
Check the following electrical connections before
you contact your Endress+Hauser service
organization:
• Verify that the sensor signal cable connector is
correctly plugged into the amplifier board
• Remote version: Check sensor and transmitter
terminal connections No. 4, 5, 6, 7.
386
S: OUTL.SENS.DEF.
$: # 386
$: # 386
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
Outlet Sensor
Defect
Defect
387
S: SEN.ASY.EXCEED
$: # 387
$: # 387
0x10
0x11
0x12
0x11
0x12
BAD
Sensor Failure
O.K.
Low
High
Sensor Asymmetry
Exceeded
Exceeded
388
S: AMP. FAULT
$: # 388
$: # 388
0x0F
BAD
Device Failure
Constant
Amplifier failure
Cause:
Amplifier error
Amplifier error
Remedy:
Contact your Endress+Hauser service organization.
Contact your Endress+Hauser service organization.
389
S: AMP. FAULT
$: # 389
$: # 389
0x0F
BAD
Device Failure
Constant
Amplifier failure
390
S: AMP. FAULT
$: # 390
$: # 390
0x0F
BAD
Device Failure
Constant
Amplifier failure
No.
Device status message
(local display)
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
diagnostic
message in the
PROFIBUS
master
Cause/remedy (spare part
→ Page 80 ff.)
Q
u
a
lit
y co
de (H
EX
)
M
e
asured va
lue st
atus
Qua
li
ty st
a
tus
Qu
ality
substatus
Li
m
its