Apex 1 to 5.5 kW Generator
Header
The first byte in each packet contains two pieces of information: five bits contain the 
packet address, and three bits contain the data byte count. If the message packet 
originates with the host computer (master), the address specifies the packet’s 
destination (to an Apex generator, for example). If the packet is going to the host, the 
address specifies the packet’s origin (from the Apex generator). The address section of 
the Header field is five bits long (bits 3-7), which allows a total of 32 distinct 
addresses. Address 0 (zero) is reserved for the network broadcast address; when this 
address is used in a host-originated packet, all units execute the packet (but do not 
respond back to the host).
The remaining three bits (bits 0, 1, and 2) tell the receiving unit how long the Data 
field is so that the unit can determine when the entire message has been received. 
Note: The value in these bits should refer only to the number of actual data bytes. Do 
not include the checksum byte when calculating the value for these bits (see 
“Checksum” on page 4-42).
Command
This field contains a one-byte value: 00h to FFh (0 to 255). If the message packet 
originates with the host computer, this value specifies the purpose of the message 
packet. If the message originates with the Apex generator, the value specifies the 
command to which it is responding. See “Host Port Commands for RS-232 with AE 
Bus” on page 4-45 f
or a complete list of commands. 
Optional Length Byte
This field supplements the Header field and exists only when the length bits in the 
Header field contain a value of 7. Under those circumstances, the Optional field 
contains a one-byte value (between 0 and 255) indicating the number of data bytes.
Data (Data Bytes)
The Data field can contain from 0 to 255 bytes of binary data, which are interpreted in 
various ways, depending on the value that appears in the Command field. The Data 
field typically contains data or a Command Status Response (CSR) (see “Creating an 
Ideal Communications Transaction” on page 4-42)
, depending on what was requested. 
Since some commands do not require data, sometimes the Data field is not present.
If the value specified in the length bits of the Header field is 0 to 6, the Apex generator 
expects 0 to 6 bytes of data. However, if the value in the Header field is 7, the Apex 
generator looks for an additional eight-bit byte after the Command field (the Optional 
field) and uses this value for the data byte count.