Bacharach Hi Flow Sampler Manuel Du Propriétaire

Page de 72
Maintenance 
Hi Flow Sampler 
 
Instruction 0055-9017 
52 
5.4 
Replacing the Internal Filters 
 
Three internal filters protect the instrument from being contaminated by 
dust and dirt. Replace these filters when they look contaminated. A good 
indication that a filter needs replaced is when the instrument’s gas-
response time increases.  
 
 
5.4.1 Leak-Gas 
Filter 
 
The leak-gas filter is located between the sampling port on the flow-
measurement tube and its associated solenoid valve on the printed 
circuit board. See Figure 5-3. 
 
Equipment Required: 
 
  Medium Phillips Head Screwdriver 
  Replacement Filter (P/N 07-1563) 
 
Procedure: 
 
1.  Remove cover per Section 5.3. 
 
2.  Remove tubing from old filter’s inlet and outlet.  
 
3.  Install tubing on new filter, making sure that the filter’s flow arrow 
points away from the flow-measurement tube. 
 
4.  Replace cover that was removed in Step 1. 
 
Figure 5-3. Leak-Gas Filter