Troubleshooting 
Hi Flow Sampler 
 
Instruction 0055-9017 
62 
TABLE 6-1. TROUBLESHOOTING GUIDE (Cont.) 
 
Instrument won’t respond when a control 
unit key is pressed. 
Microprocessor needs reset. Disconnect 
battery pack for several seconds. 
Low flow rate reading with fully charged 
battery. 
Flow path obstructed. Check flow path and 
verify flow rate per Section 5.7. 
Slow response time. 
Internal filters are dirty. Replace filters per 
Section 5.2. 
Battery power does not last at least 
4.5 hours. 
Battery pack defective. Replace. 
Erratic gas readings. 
a. The entire gas leak is not being captured. 
Use the proper attachment to capture leak. 
 
b. Faulty sensor(s). Calibrate both CH
4
 
sensors per Section 4. Replace any sensor 
that fails to calibrate. 
Sensor does not calibrate properly. 
a. Wrong calibration gas or insufficient flow 
being applied to sensor. Ensure that the 
calibration setup is correct. 
 
b. Depleted sensor. Replace sensor and 
recalibrate. 
Calibration Failed” message appears at 
bottom of display. 
Calibration parameters not within 
programmed limits due to either insufficient 
calibration gas flow, or new calibration value 
was greater than 50% of the old value. Check 
calibration setup and re-perform the 
calibration procedure. Replace sensor if 
necessary. 
Memory Full” message appears at bottom 
of display. 
1000 test records have already been stored in 
memory. Save all test records per Section 
3.15.1 and/or erase memory per Section 
3.15.3. 
No Records Saved” message appears at 
bottom of display. 
A send records command was given with no 
records stored in memory. 
Check Background Pump” message 
appears at bottom of display. 
Flow to the background-gas pump has been 
severely reduced. Check that the background 
hose is not kinked or blocked, or the internal 
tubing has not become blocked (see Figure 
5-10). Also check the condition of 
background-gas sensor filter, replace as 
necessary per Section 5.4.3. 
Check Leak Pump” message appears at 
bottom of display. 
Flow to the leak-gas pump has been severely 
reduced. Check that the internal tubing has 
not become blocked (see Figure 5-10). Also 
check the condition of leak-gas sensor filter, 
replace as necessary per Section 5.4.2. 
Check Background Calibration” message 
appears at bottom of display. 
The background-gas sensor needs calibrated 
per Section 4. 
Check Leak Calibration” message appears 
at bottom of display. 
The leak-gas sensor needs calibrated per 
Section 4.