Bacharach Hi Flow Sampler Manuel Du Propriétaire

Page de 72
Hi Flow Sampler 
Technical Data 
Instruction 0055-9017 
5
2 Technical Data 
 
Information Displayed .......... - Date and Time 
-  Battery voltage  
-  Sampling flow rate in cfm 
-  Background gas concentration in ppm or % by volume 
-  Leak concentration in ppm or % by volume  
-  Leak rate in cfm 
-  Percent difference between leak rate measurements #1 and #2 
-  Latitude, Longitude, Altitude, and Time of Fix, when the
  optional GPS receiver is attached 
 
  
Display................................. 8 line by 20 character LCD 
 
Pushbutton Controls ............ - I/O 
   (Enter) 
  (Up 
Arrow) 

(Down Arrow) 
ESC   (Escape) 
 
Communication.................... Three DB9 connectors providing serial data transfer at 115200 
baud to a personal computer, bar code reader, and GPS receiver 
 
Measured Values.................. - Sampling flow rate 
- Background gas concentration 
- Sample gas concentration  
- Battery voltage 
 
Calculated Values ................ - Leak concentration corrected for background gas level 
- Leak rate 
- Percent difference between leak rate measurements #1 and #2 
 
Measurable Leak Rate .......... 0.05 to 8.00 SCFM (1.42 to 226 LPM) 
0.05 to 6.00 SCFM (1.42 to 170 LPM) 
 
Accuracy of Calculated  
Leak Rate ............................. ±10% of reading 
 
Temperature: 
  Operating .......................... 0 to 50 °C (32 to 122 °F) 
  Storage.............................. –40 to 60 °C (–40 to 140 °F) 
 
Humidity.............................. 5 to 95% RH (non-condensing) 
 
Sampling Flow Rate: 
  Maximum .......................... 10.5 SCFM (297 LPM) at full battery charge 
  Operating Flow Points ....... Initial flow approx. 10 SCFM (283 LPM) 
Second flow approx. 8 SCFM (226 LPM) 
(The second flow rate is 25% of the initial flow) 
  Measurement Method........ Differential pressure across restriction 
 Accuracy ........................... ±5% 
of reading