Equipment not complying with IEC 60601 must be kept outside the 
patient environment, as defined in the standard as at least 1.5 meters 
from the patient or the patient support. 
Any person who connects external equipment to signal input, signal 
output, or other connectors has formed a system and is therefore 
responsible for the system to comply with the requirements of IEC 
60601-1-1. If in doubt, speak with a qualified technician. 
In locations where 240V outlets are used, connect this display only on a 
center-tapped, 240V, single-phase supply. 
Water and moisture   
Never expose the display to rain or moisture.   
Never use the display near water - e.g. near a bathtub, washbasin, 
swimming pool, kitchen sink, laundry tub or in a wet basement. 
Ventilation 
Do not cover or block the ventilation openings in the cover of the set. 
When installing the display in a cupboard or another closed location, 
heed the necessary space between the set and the sides of the 
cupboard. 
Installation  
Place the display on a flat, solid and stable surface that can support the 
weight of at least 3 displays. If you use an unstable cart or stand, the 
display may fall, causing serious injury to a child or adult, and serious 
damage to the equipment.