MR810 Technical Manual Rev C  REF 185042601
 
 
 
3 
3.5.2 N
ON 
H
EATER
-
WIRE 
M
ODE
 
If no heater-wire is detected, the MR810 initiates non heater-wire mode.  In this mode the heater plate 
temperature is controlled to a fixed value.  Heater plate temperatures for each temperature setting are as 
follows: 
Low Setting: 
45 °C 
Medium Setting:  60 °C 
High Setting: 
70 °C 
As this mode controls to a fixed heater plate temperature, any variation in ambient temperature has no effect 
on the control algorithm. 
3.6 D
IAGNOSTIC PORT
 
The diagnostic port is used for downloading information when servicing of the unit is required.  It must be 
used with the communication cable found in the service kit REF 043042263 to connect to a PC. 
 
 
Figure 3.5 Diagnostic port