MR810 Technical Manual Revision C  REF 185042601
 
 
2
2 S
PECIFICATIONS
 
2.1 M
ECHANICAL
 
Dimensions: 
94 x 154 x 135 mm (without chamber fitted). 
Weight: 
1.7 kg without chamber. 
 
2.0 kg with MR290 filled with water. 
2.2 E
LECTRICAL
 
Supply Voltage: 
230 V~ (MR810A--) 
115 V~ (MR810J--) 
100 V~ (MR810G--) 
Supply Frequency: 
50 or 60 Hz 
Supply Current: 
0.8 A maximum at 230 V~ (MR810A--) 
1.6 A maximum at 115 V~ (MR810J--) 
1.8 A maximum at 100 V~ (MR810G--) 
Heater-wire: 
22 V~ at 30 W 
Heater Plate: 
150 W 
Heater Plate Thermal Cutout: 
Operates at 93 ± 5 
°
 
2.3 P
ERFORMANCE
 
Note: Performance results obtained at 23 °C ambient temperature. 
2.3.1 H
EATER
-
WIRE 
M
ODE 
P
ERFORMANCE
 
Table 1: Heater-wire mode performance for each temperature setting 
Temperature 
Setting 
Constant flow range 
(L/min) 
Delivered patient temperature (°C) 
Low 
5 to 60 
26 to 29 
Medium 
5 to 60 
30 to 33 
High 
5 to 30 
33 to 36 
Results obtained using an RT308 oxygen therapy breathing circuit. 
2.3.2 N
ON 
H
EATER
-
WIRE 
M
ODE 
P
ERFORMANCE
 
Table 2: Non heater-wire mode performance for each Temperature Setting 
Temperature 
Setting 
Constant flow range 
(L/min) 
Delivered patient 
temperature (°C) 
Low 
5 to 60 
23 to 25 
Medium 
5 to 40 
25 to 27 
High 
5 to 30 
28 to 32 
Results obtained using an RT307 oxygen therapy breathing circuit.