Fisher & Paykel MR810 Manuel D’Utilisation
MR810 Technical Manual Revision C REF 185042601
7
2
2 S
PECIFICATIONS
2.1 M
ECHANICAL
Dimensions:
94 x 154 x 135 mm (without chamber fitted).
Weight:
1.7 kg without chamber.
2.0 kg with MR290 filled with water.
2.2 E
LECTRICAL
Supply Voltage:
230 V~ (MR810A--)
115 V~ (MR810J--)
100 V~ (MR810G--)
115 V~ (MR810J--)
100 V~ (MR810G--)
Supply Frequency:
50 or 60 Hz
Supply Current:
0.8 A maximum at 230 V~ (MR810A--)
1.6 A maximum at 115 V~ (MR810J--)
1.8 A maximum at 100 V~ (MR810G--)
1.6 A maximum at 115 V~ (MR810J--)
1.8 A maximum at 100 V~ (MR810G--)
Heater-wire:
22 V~ at 30 W
Heater Plate:
150 W
Heater Plate Thermal Cutout:
Operates at 93 ± 5
°
C
2.3 P
ERFORMANCE
Note: Performance results obtained at 23 °C ambient temperature.
2.3.1 H
EATER
-
WIRE
M
ODE
P
ERFORMANCE
Table 1: Heater-wire mode performance for each temperature setting
Temperature
Setting
Constant flow range
(L/min)
Delivered patient temperature (°C)
Low
5 to 60
26 to 29
Medium
5 to 60
30 to 33
High
5 to 30
33 to 36
Results obtained using an RT308 oxygen therapy breathing circuit.
2.3.2 N
ON
H
EATER
-
WIRE
M
ODE
P
ERFORMANCE
Table 2: Non heater-wire mode performance for each Temperature Setting
Temperature
Setting
Constant flow range
(L/min)
Delivered patient
temperature (°C)
temperature (°C)
Low
5 to 60
23 to 25
Medium
5 to 40
25 to 27
High
5 to 30
28 to 32
Results obtained using an RT307 oxygen therapy breathing circuit.