Status alarms and Status Alarm Severity
 (continued)
Table 5-2:   
Alarm code
Status message
Default severity Notes
Configurable?
A114
mA Output 2 Fixed
Informational
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
A115
No External Input or Polled
Data
Informational
Yes
A116
Temperature Overrange
(Petroleum)
Informational
Applies only to transmitters with
the petroleum measurement appli-
cation.
Yes
A117
Density Overrange (Petro-
leum)
Informational
Applies only to transmitters with
the petroleum measurement appli-
cation.
Yes
A118
Discrete Output 1 Fixed
Informational
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
A119
Discrete Output 2 Fixed
Informational
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
A120
Curve Fit Failure (Concen-
tration)
Informational
Applies only to transmitters with
the concentration measurement
application.
No
A121
Extrapolation Alarm (Con-
centration)
Informational
Applies only to transmitters with
the concentration measurement
application.
Yes
A131
Meter Verification in Pro-
gress: Outputs to Last
Measured Value
Informational
Applies only to transmitters with
Smart Meter Verification.
Yes
A132
Sensor Simulation Active
Informational
Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
A141
DDC trigger(s) have com-
pleted
Informational
Applies only to flowmeters with the
enhanced core processor.
Can be set to either Informational or
Ignore
, but cannot be set to Fault.
Yes
5.6
Configure informational parameters
The informational parameters can be used to identify or describe your flowmeter but they
are not used in transmitter processing and are not required.
The informational parameters include:
Device parameters
-
Descriptor
Configure device options and preferences
94
Micro Motion
®
 Model 2700 Transmitters with Analog Outputs