Typical discrete output circuit
Figure 6-1:   
A.
24  V (Nom)
B.
3.2  K
Ω
C.
Out+
D.
Out
6.4.3
Configure Discrete Output Fault Action
Display
Not available
ProLink II
ProLink
 > Configuration > Frequency/Discrete Output > Discrete Output > DO Fault Action
ProLink III
Device Tools
 > Configuration > Fault Processing
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Fault Action
Overview
Discrete Output Fault Action
 controls the behavior of the discrete output if the transmitter
encounters an internal fault condition.
Note
For some faults only: If Last Measured Value Timeout is set to a non-zero value, the transmitter will not
implement the fault action until the timeout has elapsed.
CAUTION!
Do not use Discrete Output Fault Action as a fault indicator. If you do, you may not be able to
distinguish a fault condition from a normal operating condition. If you want to use the discrete
output as a fault indicator, see 
Integrate the meter with the control system
Configuration and Use Manual 