Emerson MMI-20019043 Manuel D’Utilisation

Page de 374
Overview
Totalizers keep track of the total amount of mass or volume measured by the transmitter
since the last totalizer reset. Inventories keep track of the total amount of mass or volume
measured by the transmitter since the last inventory reset.
Tip
You can use the inventories to keep a running total of mass or volume across multiple totalizer
resets.
9.6
Start and stop totalizers and inventories
Display
See 
.
ProLink II
ProLink
 > Totalizer Control > Start
ProLink
 > Totalizer Control > Stop
ProLink III
Device Tools
 > Totalizer Control > Totalizer and Inventories > Start All Totals
Device Tools
 > Totalizer Control > Totalizer and Inventories > Stop All Totals
Field Communicator Service Tools > Variables > Totalizer Control > All Totalizers > Start Totalizers
Service Tools
 > Variables > Totalizer Control > All Totalizers > Stop Totalizers
Overview
When you start a totalizer, it tracks process measurement. In a typical application, its value
increases with flow. When you stop a totalizer, it stops tracking process measurement and
its value does not change with flow. Inventories are started and stopped automatically,
when totalizers are started and stopped.
Important
Totalizers and inventories are started or stopped as a group. When you start any totalizer, all other
totalizers and all inventories are started simultaneously.When you stop any totalizer, all other
totalizers and all inventories are stopped simultaneously. You cannot start or stop inventories
directly.
9.6.1
Start and stop totalizers and inventories using the
display
Prerequisites
The Totalizer Start/Stop display function must be enabled.
At least one totalizer must be configured as a display variable.
Procedure
To start all totalizers and inventories using the display:
Transmitter operation
158
Micro Motion
®
 Model 2700 Transmitters with Analog Outputs