The controls are not accessible in any other way.
10.2.1
Switch between secured and unsecured mode using
ProLink II
Prerequisites
Before switching to unsecured mode, ensure that you will be able to switch back to
secured mode. Because switching to unsecured mode requires breaking the physical seal,
switching back to secured mode may require a site visit from a certified Weights &
Measures inspector and reinstallation of the physical seal.
Ensure that the wiring is in place for a service port connection. You may need to break the
seal and remove the clamp.
Procedure
To switch from secured mode to unsecured mode:
1. Make a service port connection to your transmitter.
2. Choose Plug-ins > Enable/Disable Custody Transfer.
3. Click Disable Custody Transfer.
To switch from unsecured mode to secured mode:
1. Make a service port connection to your transmitter.
2. Choose Plug-ins > Enable/Disable Custody Transfer
3. Click Enable Custody Transfer.
10.2.2
Switch between secured and unsecured mode using
ProLink III
Prerequisites
Before switching to unsecured mode, ensure that you will be able to switch back to
secured mode. Because switching to unsecured mode requires breaking the physical seal,
switching back to secured mode may require a site visit from a certified Weights &
Measures inspector and reinstallation of the physical seal.
Ensure that the wiring is in place for a service port connection. You may need to break the
seal and remove the clamp.
Procedure
To switch from secured mode to unsecured mode:
1. Make a service port connection to your transmitter.
2. Choose Device Tools > Configuration > Weights & Measures.
3. Set Software Security to Disabled.
To switch from unsecured mode to secured mode:
Operate the transmitter with the Weights & Measures application
Configuration and Use Manual 