Emerson MMI-20019043 Manuel D’Utilisation

Page de 374
Smart Meter Verification flowchart: Scheduling test execution using the display
Scheduling Smart Meter Verification test execution using the display
Figure 11-6:   
SCHEDULE VERFY
Select
SAVE/YES?
TURN OFF SCHED/YES?
SET NEXT
Select
Select
Schedule set?
Yes
Schedule deleted
Scroll
HOURS LEFT
Select
Scroll
xx HOURS
SET RECUR
No
SCHED IS OFF
xx HOURS
Select
Scroll
No
Yes
Select
xx HOURS
Select
SAVE/YES?
Scroll
No
Yes
Scroll
Scroll
EXIT
Scroll
Select
Select
Scroll
Manage scheduled test execution using ProLink II
1.
Choose Tools > Meter Verification > Schedule Meter Verification.
2.
To schedule a single test or the first test in recurring execution, specify a value for
Hours Until Next Run
.
3.
To schedule recurring execution, specify a value for Hours Between Recurring Runs.
Measurement support
Configuration and Use Manual 
185