Emerson MMI-20019043 Manuel D’Utilisation

Page de 374
D1 and D2 density calibration using ProLink III
Figure 11-8:   
Enter density of D1 fluid
D1 calibration
Close shutoff valve 
downstream from sensor
Fill sensor with D1 fluid
Done
Device Tools >
  Calibration >
     Density Calibration >
        Density Calibration – Point 1 (Air)
Close
Start Calibration
Enter density of D2 fluid
D2 calibration
Fill sensor with D2 fluid
Device Tools >
  Calibration >
     Density Calibration >
        Density Calibration – Point 2 (Water)
Close
Start Calibration
Postrequisites
If you disabled LD Optimization before the calibration procedure, re-enable it.
11.5.3
Perform a D1 and D2 density calibration using the
Field Communicator
Prerequisites
During density calibration, the sensor must be completely filled with the calibration
fluid, and flow through the sensor must be at the lowest rate allowed by your
application. This is usually accomplished by closing the shutoff valve downstream
from the sensor, then filling the sensor with the appropriate fluid.
D1 and D2 density calibration require a D1 (low-density) fluid and a D2 (high-
density) fluid. You may use air and water.
If LD Optimization is enabled on your meter, disable it. To do this, choose Configure >
Manual Setup
 > Measurements > LD Optimization. LD Optimization is used only with large
sensors in hydrocarbon applications. In some installations, only Micro Motion
customer service has access to this parameter. If this is the case, contact
Micro Motion before continuing.
The calibrations must be performed without interruption, in the order shown. Make
sure that you are prepared to complete the process without interruption.
Measurement support
Configuration and Use Manual 
197