Emerson MMI-20019043 Manuel D’Utilisation

Page de 374
D3 or D3 and D4 density calibration using the Field Communicator
Figure 11-12:   
Enter density of D3 fluid
Density Calibration 
Complete message
Calibration in Progress 
message
D3 calibration
Close shutoff valve 
downstream from sensor
Fill sensor with D3 fluid
Fill sensor with D4 fluid
D4 calibration
OK
Dens Pt 3 T-Series
Calibration method 
executes
OK
Home
Enter density of D4 fluid
Density Calibration 
Complete message
Calibration in Progress 
message
OK
Dens Pt 4 T-Series
Calibration method 
executes
OK
Home
Done
On-Line Menu >
     Service Tools >
          Maintenance >
               Density Calibration
Service Tools >
          Maintenance >
               Density Calibration
Done
11.7
Perform temperature calibration
Temperature calibration establishes the relationship between the temperature of the
calibration fluids and the signal produced by the sensor.
Prerequisites
The temperature calibration is a two-part procedure: temperature offset calibration and
temperature slope calibration. The two parts must be performed without interruption, in
the order shown. Make sure that you are prepared to complete the process without
interruption.
You must use ProLink II or ProLink III to calibrate for temperature.
Measurement support
Configuration and Use Manual 
203