Emerson MMI-20019043 Manuel D’Utilisation

Page de 374
12.3
Flow measurement problems
Flow measurement problems and recommended actions
Table 12-3:   
Problem
Possible causes
Recommended actions
Flow indication at no
flow conditions or
zero offset
Misaligned piping (especially in new in-
stallations)
Open or leaking valve
Incorrect sensor zero
Verify that all of the characterization parame-
ters match the data on the sensor tag.
If the flow reading is not excessively high, re-
view the live zero. You may need to restore
the factory zero.
Check for open or leaking valves or seals.
Check for mounting stress on the sensor
(e.g., sensor being used to support piping,
misaligned piping).
Contact Micro Motion.
Erratic non-zero flow
rate under no-flow
conditions
Leaking valve or seal
Slug flow
Plugged or coated flow tube
Incorrect sensor orientation
Wiring problem
Vibration in pipeline at rate close to sen-
sor tube frequency
Damping value too low
Mounting stress on sensor
Verify that the sensor orientation is appropri-
ate for your application (refer to the sensor
installation manual).
Check the drive gain and the pickoff voltage.
See 
If the wiring between the sensor and the
transmitter includes a 9-wire segment, verify
that the 9-wire cable shields are correctly
grounded.
Check the wiring between the sensor and
transmitter. See 
.
For sensors with a junction box, check for
moisture in the junction box.
Purge the flow tubes.
Check for open or leaking valves or seals.
Check for sources of vibration.
Verify damping configuration.
Verify that the measurement units are con-
figured correctly for your application.
Check for slug flow. See 
Check for radio frequency interference. See 
.
Contact Micro Motion.
Troubleshooting
220
Micro Motion
®
 Model 2700 Transmitters with Analog Outputs