Restrictions
This parameter does not apply to inventories. You cannot reset inventories from the display.
You cannot use the display to reset all totalizers as a group. You must reset totalizers
individually.
If the petroleum measurement application is installed on your computer, the operator must
enter the off-line password to perform this function, even if the off-line password is not
enabled.
Procedure
1.
Ensure that the totalizers you want to reset have been configured as display
variables.
If the totalizer is not configured as a display variable, the operator will not be able to
reset it.
2.
Enable or disable resetting the totalizer as desired.
Option
Description
Enabled
Operators can reset a totalizer from the display, if the totalizer is config-
ured as a display variable.
Disabled
 (default)
Operators cannot reset totalizers from the display.
5.2.3
Enable or disable the Acknowledge All Alarms display
command
Display
OFF-LINE MAINT
 > OFF-LINE CONFG > DISPLAY > ALARM
ProLink II
ProLink
 > Configuration > Display > Display Options > Display Ack All Alarms
ProLink III
Device Tools
 > Configuration > Transmitter Display > Ack All
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features > Acknowledge All
Overview
You can configure whether or not the operator can use a single command to acknowledge
all alarms from the display.
Procedure
1.
Ensure that the alarm menu is accessible from the display.
To acknowledge alarms from the display, operators must have access to the alarm
menu.
2.
Enable or disable Acknowledge All Alarms as desired.
Configure device options and preferences
84