Emerson 3000 Manuel D’Utilisation

Page de 76
Instruction Bulletin
106-300NE Rev. 3.4
May 2000
Rosemount Analytical Inc.    A Division of Emerson Process Management 
Installation     2-7
World Class 3000
VEE
DEFLECTOR
VEE
DEFLECTOR
DIFFUSION
ELEMENT
SETSCREW
FILTER
GAS FLOW
DIRECTION
APEX
P0017
Figure 2-2.  Orienting the Optional Vee Deflector
c. Service 
Required
 
1.  Power input: 100, 115 or 220 Vac sin-
gle phase, 50 to 60 Hz, 3 amp
minimum. (See label.)
 
2.  Compressed air: 10 psig (68.95 kPag)
minimum, 225 psig (1551.38 kPag)
maximum at 2 scfh (56.6 L/hr) maxi-
mum; supplied by one of the following
(less than 40 parts-per-million total hy-
drocarbons). Regulator outlet pressure
should be set at 5 psi (35 kPa).
(a)  Instrument air - clean, dry.
(b)  Bottled standard air with step-down
regulator.
(c)  Bottled compressed gas mixture
(20.95% oxygen in nitrogen).
(d)  Other equivalent clean, dry, oil-free
air supply.
Figure 2-3.  Outline of Heater Power Supply
2