Emerson 1056 Manuel D’Utilisation

Page de 86
21
MODEL FCL-1056
SECTION 5.0
PROGRAMMING THE ANALYZER
TABLE 5-1. DEFAULT SETTINGS (continued)
ITEM
CHOICES
DEFAULT
Measurement
1.   Chlorine (sensor 1)
a.   units
ppm or mg/L
ppm
b.   resolution
0.01 or 0.001
0.001
c.   input filter
0 to 999 sec
5 sec
2.   pH (sensor 2)
a.   pre-amplifier location
analyzer or sensor/junction box
analyzer
b.   solution temperature correction
on or off
off
c.   resolution
0.01 or 0.1
0.01
d.   input filter 
0 to 999 sec
2 sec
e.   reference impedance
low or high
low
Temperature related settings
1.   Units
ºC or ºF
ºC
2.   Temperature compensation
automatic or manual
automatic
Security Code
1.   Calibrate/Hold
000 to 999
000
2.   Program/Display
000 to 999
000
pH Sensor Diagnostic Limits
1.   Reference offset
0 to 9999 mV
60 mV
2.   Diagnostics
on or off
on
3.   Glass impedance temperature 
correction
on or off
on
4.   Glass fault (low impedance)
0 to 9999 M
Ω
10 M
Ω
5.   Glass fault (high impedance)
0 to 9999 M
Ω
1500 M
Ω
6.   Reference fault (high impedance)
0 to 9999 k
Ω
40 k
Ω
Calibration–pH
1.   Stabilization criteria
a.   time interval
0 to 99 sec
10 sec
b.   pH change
0.01 to 1.00 pH
0.02 pH
2.   User-entered slope
0.00 to 99.99 mV/pH
59.16 mV/pH
3.   User-entered offset
-999 to +999 mV
0 mV
Calibration–Analog Outputs
1.   4 mA
0.000 to 22.000 mA
4.000 mA
2.   20 mA
0.000 to 22.000 mA
20.000 mA