Emerson 1056 Manuel D’Utilisation

Page de 86
MODEL FCL-1056                                                                                                                                            
SECTION 9.0
TROUBLESHOOTING
Fault message
Explanation
Section
Main Board CPU Error
Main board software is corrupted
9.3.1
Main Board Factory Data
Main board factory eeprom data is corrupted
9.3.1
Main Board User Data
Main board user eeprom data is corrupted
9.3.1
Sensor Hardware Error
Missing or bad hardware component
9.3.2
Sensor Board Unknown
Analyzer does not recognize sensor board
9.3.3
Sensor HW-SW Mismatch
Sensor board hardware and software do not match
9.3.3
Sensor Incompatible
Sensor board software is not supported by main board     
software
9.3.4
Sensor Not Communicating
Sensor board is not communicating with main board
9.3.3
Sensor CPU Error
Sensor board software is corrupted
9.3.5
Sensor RTD Open
Temperature measuring circuit is open 
9.3.6
S1 Not Detected
No sensor board is connected to sensor 1 terminal
9.3.7
Sensor Factory Data
Sensor board factory eeprom data is corrupted
9.3.8
Sensor EEPROM Write Error
Bad CPU on the sensor board
9.3.8
Sensor User Data
Sensor board user eeprom data is corrupted
9.3.8
Sensor ADC Error
Bad component on the sensor board
9.3.9
Sensor RTD Out of Range
RTD is improperly wired or has failed
9.3.10
Sensor Glass Z Too High
The impedance of the pH sensing glass membrane is   
too high.
9.3.11
Sensor Glass Z Too High
The impedance of the pH reference function is 
too high.
9.3.12
Sensor Broken Glass
The impedance of the pH sensing glass membrane is                                 
very low, suggesting a broken glass membrane
9.3.13
9.3 TROUBLESHOOTING WHEN A FAULT MESSAGE IS SHOWING
9.3.1 Main Board CPU, Main Board Factory Data, and Main Board User Data Errors.
These error messages mean the main board software is corrupted or the eeprom data on the main board is
corrupted. 
1.    Cycle the power off then on.
2.    If cycling the power does not help, call the factory. The main board must be replaced. To do this, the analyzer
must be returned to the factory.
3.    If cycling the power does not help and the fault message was Main Board User Data, reset the analyzer to 
factory default and re-enter user settings and repeat calibration.
9.3.2 Hardware Error.
Hardware error means there is a missing or bad hardware component on the sensor board. The board must be
replaced.
9.3.3 Sensor Board Unknown, Sensor Board HW (Hardware) or SW (Software) Mismatch, or Sensor Board
Not Communicating.
These error messages mean the main board either does not recognize the sensor board or the sensor board and
main board are no longer communicating. 
1.    Verify that the ribbon cable connecting the main board (on the inside of the front panel) and the sensor board    
are properly seated. Inspect the connecting cable for obvious tears or breaks.
2.    If the ribbon cable is properly seated and appears undamaged, the sensor board should be replaced. 
60