GE CTR 80 Manuel D’Utilisation

Page de 61
50
10 Digital Communication Interface
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Display Temperature
Read current set-point
s[etpoint]
s
set: 999.99 {C or F}
set: 90.00 C
Set current set-point to n
s[etpoint]=n
s=80.00
Instrument
Range
Read temperature
t[emperature]
t
t: 999.99 {C or F}
t: 55.6 C
Read temperature units
u[nits]
u
u: x
u: C
C or F
Set temperature units to Celsius
u[nits]=c
u=c
Set temperature units to
Fahrenheit
u[nits]=f
u=f
Read scan mode
sc[an]
sc
scan: {ON or OFF}
scan:ON
Set scan mode
sc[an]=on/off
sc=on
ON or OFF
Read scan rate
sr[ate]
sr
srat: 99.9 {C or F}/min
srat:12.4C/min
Set scan rate
sr[ate]=n
sr=1.1
.1 to 99.9
Secondary Menu
Read proportional band setting
pr[opband]
pr
pb: 999.9
pb: 15.9
Set proportional band to
n
pr[opband]=n
pr=8.83
Depends on
Configuration
Read cut-out setting
cu[tout]
cu
cu:9999 {C or F}
cu:105 C, in
Set cut-out setting to
degrees
cu[tout]=
n
cu-95
25 to 115°C
Read heater power
(duty cycle)
po[wer]
po
po: 999.9
po: 1.0
Configuration Menu
Read high limit
hl
hl
hl:999
hl:126
Set high limit
hl=n
hl=90
25 to 100
Read low limit
ll
ll
ll:999
ll:-90
Set low limit
ll=n
ll=-90
–90 to 25
Read cooling setting
co[ol]
co
cool: {ON or OFF}
cool: ON
Set cooling setting
co[ol]=on/of[f]
co=on
ON or OFF
Read serial sample setting
sa[mple]
sa
sa: 9
sa: 1
Set serial sampling setting to
n
seconds
sa[mple]=n
sa=0
0 to 999
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Table 3
Controller Communications Commands