HIGH-LEVEL DISINFECTION
RINSING
TRANSPORTATION
DRYING
NOMENCLATURE
IMPORTANT
1
2
3
4
5
6
9
10
7
8
Attach a syringe to the adapter. 
Pull the plunger of the syringe to
fill the instrument/suction channels
and channel cleaning adapter
with disinfectant solution.
Fill basin with disinfectant. Attach a 
suction cleaning adapter (MAJ-222) to 
the previously cleaned bronchoscope. 
Immerse completely in disinfectant 
solution.
While immersed, detach the adapter 
from the bronchoscope. Use a clean, 
lint-free cloth to wipe away any bub-
bles clinging to external surfaces.
Soak all equipment according 
to the time and temperature 
recommended by the disinfectant 
manufacturer. Use a timer to mea-
sure the exact contact time.
Connect the suction cleaning adapter 
and syringe. Inject air to remove 
disinfectant from the instrument/suction 
channels, then remove the broncho-
scope from the disinfectant. 
Disconnect the suction cleaning 
adapter and immerse the broncho-
scope and adapter in sterile water. 
Using a sterile, lint-free cloth, thor-
oughly rinse and wipe all surfaces.
Connect suction cleaning adapter
to bronchoscope and to a tube 
from a suction pump. Aspirate 
water through channels for 30 
seconds. Remove from water and 
aspirate air through channels for 
60 seconds.
Turn the suction pump off. Hold the control 
section with the openings pointing down and 
disconnect the suction cleaning adapter.
Dry the bronchoscope and adapter with an 
alcohol-moistened cloth. Make sure that the 
equipment is completely dry prior to storage.
Hold all parts of the endoscope when carrying. Do 
not let the ends of the bronchoscope swing freely.
Clean the bronchoscope 
meticulously prior to disinfection.
• Wear appropriate personal protective 
equipment during reprocessing.
• Autoclave all autoclavable accessories.
• Store the bronchoscope hanging vertically 
with all valves and caps removed.
• Use a compatible disinfectant and test it
for proper concentration as recommended 
by the manufacturer.
• Do not use the carrying case to store 
"patient-ready" endoscopes. Any 
bronchoscope placed in the carrying case 
must be reprocessed prior to patient use.
Immerse the bronchoscope's distal 
end in 70% alcohol and aspirate 
alcohol for 5 seconds. Remove the 
distal end from the alcohol and
aspirate air for 20 seconds.
©
 2
0
0
4
 O
ly
m
pu
A
m
er
ic
In
c.
 /
 P
ri
nt
ed
 in
 U
SA
  
0
1
/0
4
 —
 P
W
C
-1
0
0
0
2
0
 for OES
®
 BF-40 Series Bronchofiberscopes
BRONCHOSCOPE DISINFECTION GUIDE
 Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information
WARNING:
 
This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the 
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.