ROUBLESHOOTING 
G
UIDE
 
 
SYMPTOMS POSSIBLE 
CAUSE(S) 
SOLUTION 
The host is 
receiving data but 
the data does not 
look correct. 
The scanner and host 
may not be configured 
for the same interface. 
Check that the scanner and the 
host are configured for the same 
interface. 
Characters are 
being dropped. 
Intercharacter delay 
needs to be added to 
the transmitted output. 
Add some intercharacter delay to 
the transmitted output by using the
MetroSelect Configuration Guide 
(
MLPN
 00-02407). 
Aux Port Operation with any Interface 
Trouble with the 
secondary 
scanner. 
 
Refer to the user guide provided 
with the secondary scanner. 
Cable [
MLPN
 54-54667] 
may not be connected 
to the proper port. 
Ensure the secondary scanner is 
connected to the MS7320 com 
port marked “Aux” port. 
*  The MS7320 must be 
configured to enable the “Aux” 
port. 
Secondary 
scanner powers 
up but data is 
not relayed to 
the host. 
The “Aux” com port 
may not be operating 
properly. 
The secondary scanner must be 
configured to send ‘secondary’ 
formatted data (reserve  
code 32). 
*   Use MetroSet
®
. For the Auxiliary interface, choose “HoloTrak Decode”.  All 
remaining parameters will be automatically chosen.
  
USB Only
 
The scanner 
Powers up ok, 
scans ok but 
does not 
communicate. 
The USB Port is not 
operating correctly. 
Check that the scanner is 
configured for USB operation. 
 
Check that the host’s USB port is 
enabled. 
The scanner 
emits a razz tone 
and the LEDS 
flash three times 
when configured 
for USB.  
The USB port is not 
operating correctly. 
Disconnect then reconnect the 
USB cable at the host end. 
Contact a Metrologic  
Representative if symptoms 
persist.