Calibration 
Series N5700 User’s Guide 
111 
Calibration 
Refer to the “Equipment Required” section in this appendix for a list 
of the equipment required for calibration. A general outline of the 
procedure is as follows: 
 
Enter the calibration mode by providing the correct password. 
The password is set at the factory to the model number. Once 
calibration has been entered, the password can be changed by the 
user. lf the password is incorrect, an error occurs.  
 
You do not have to do a complete calibration. If appropriate, you 
may calibrate only the voltage or current functions and then save 
the calibration constants.  
 
As each calibration sequence is completed, the instrument saves 
the calibration constants and begins using them.  
 
Exit the calibration mode. Note that a Reset command also sets 
the calibration state to OFF.
 
Calibration Procedure 
Unless instructed otherwise, connect the +sense terminal to the 
+output, and the -sense terminal to the -output. 
When calibrating the unit using SCPI commands, most calibration 
steps involve sending an *OPC? query to synchronize with the power 
supply’s command completion before proceeding. The response from 
the instrument must be read each time *OPC? is given. 
NOTE 
The CAL:LEV and CAL:DATA commands may take several seconds to complete. 
If a timeout occurs in your VISA application, you may need to change the 
VI_ATTR_TMO_VALUE in the ViSetAttribute function. 
Enable Calibration mode 
*RST 
OUTP ON 
CAL:STAT ON, <password> 
Voltage Programming and Measurement Calibration 
Step 1. Connect the Agilent 3458A voltage input to an output. 
Step 2. Select the full-scale voltage programming range. The value to program 
a range must be the maximum voltage of the range. 
CAL:VOLT <voltage> 
Step 3. Select the first voltage calibration point. 
CAL:LEV P1